EasyManua.ls Logo

Keysight M9421A - Questionable Calibration Negative Transition; Questionable Calibration Positive Transition

Keysight M9421A
2667 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Questionable Calibration Negative Transition
Determines which bits in the "Questionable Calibration Condition" on page 2622
register will set the corresponding bit in the "Questionable Calibration Event Query"
on page 2623 register when the condition register bit has a negative transition (1 to
0).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:NTRansition <integer>
:STATus:QUEStionable:CALibration:NTRansition?
Example
:STAT:QUES:CAL:NTR 16384
“Align All Now Needed” being cleared will be reported to the Calibration Summary of the Status
Questionable register
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Questionable Calibration Positive Transition
Determines which bits in the "Questionable Calibration Condition" on page 2622
register will set the corresponding bit in the "Questionable Calibration Event Query"
on page 2623 register when the condition register bit has a positive transition (0 to
1).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:PTRansition <integer>
:STATus:QUEStionable:CALibration:PTRansition?
Example
:STAT:QUES:CAL:PTR 16384
“Align All Now Needed” being set will be reported to the Calibration Summary of the Status
Questionable register
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
WLAN Mode User's &Programmer's Reference 2624

Table of Contents