EasyManua.ls Logo

Keysight M9421A - 3.12.8 Meas Setup; 3.12.8.1 Settings; Avg;Hold Number (Averaging On;Off); Average Mode

Keysight M9421A
2667 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 WLAN Mode
3.12 IQ Waveform Measurement
3.12.8 Meas Setup
Contains functions for setting up the measurement parameters, and for setting up
parameters global to all measurements in the Mode.
3.12.8.1 Settings
Contains frequently used Meas Setup functions to which you will want the fastest
access.
Avg/Hold Number (Averaging On/Off)
Sets the number of sweeps (average counts) that are averaged. After the specified
number of sweeps, the "Average Mode" on page 1825 (terminal control) setting
determines the averaging action.
Also lets you turn Averaging on or off.
Remote
Command
[:SENSe]:WAVeform:AVERage:COUNt <integer>
[:SENSe]:WAVeform:AVERage:COUNt?
Example
:WAV:AVER:COUN 1001
:WAV:AVER:COUN?
Preset 10
State Saved Saved in instrument state
Min/Max 1/20001
Annotation The average count is displayed in the measurement bar on the front panel display. The annotation
appears in the format n/N, where n is the current average and N is the average count
Auto Function
Remote Command
[:SENSe]:WAVeform:AVERage[:STATe] OFF | ON | 0 | 1
[:SENSe]:WAVeform:AVERage[:STATe]?
Example
:WAV:AVER ON
:WAV:AVER?
Preset
OFF
State Saved Saved in instrument state
Range
OFF|ON
Average Mode
Sets the Average Mode:
1825 WLAN Mode User's &Programmer's Reference