3 WLAN Mode
3.11 Monitor Spectrum Measurement
<meas> is the identifier for the current measurement; any one of CHPower-
|ACPower|OBWidth|MONitor
Example Channel Power measurement
:CHP:SWE:ETIMe?
Dependencies Available only on non-sweeping hardware
Preset Automatically calculated
3.11.9.2 Sweep Config
Accesses controls that enable you to configure the Sweep and Control functions of
the instrument, such as Sweep Rules.
Points
Sets the number of points taken per sweep, and displayed in the traces. The current
value of points is displayed parenthetically, next to the sweep time in the lower-right
corner of the display. Using more points provides greater resolution. Using fewer
points compacts the data and decreases the time required to access a trace over
the remote interface.
Increasing the number of points does not increase the sweep time. However, it can
slightly impact the trace processing time and therefore the overall measurement
speed. Decreasing the number of points does not decrease the sweep time, but it
may speed up the measurement, depending on the other sweep settings (for
example, in FFT sweeps). Fewer points will always speed up the I/O.
Due to minimum sweep rate limitations of the hardware, the minimum sweep time
available to the user will increase above its normal value of 1 ms as the number of
sweep points increases above 15001.
Changing the number of sweep points has several effects on the instrument. Since
markers are read at the point location, the marker reading may change. The sweep
time resolution will change. Trace data for all the traces will be cleared and, if
Sweep is in Cont, a new trace taken. If any trace is in average or hold, the averaging
starts over.
Because of sweep time quantization issues, the knob and up/down keys cannot be
used to adjust the number of points.
When in a split screen display each window may have its own value for points.
When sweep points is changed, an informational message "Sweep points changed,
all traces cleared" is displayed and in the 5G NR mode, Auto Sweep Points is set to
OFF(0).
Remote Command
[:SENSe]:MONitor:SWEep:POINts <integer>
WLAN Mode User's &Programmer's Reference 1726