EasyManua.ls Logo

Keysight M9421A - Averaging On;Off; Avg Mode

Keysight M9421A
2667 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 WLAN Mode
3.6 Modulation Analysis Measurement
Example
:EVM:AVER:COUN 1
:EVM:AVER:COUN?
Preset 10
State Saved Saved in instrument state
Min 1
Max 10000
Averaging On/Off
Turns averaging on or off.
Remote Command
[:SENSe]:EVM:AVERage[:STATe] OFF | ON | 0 | 1
[:SENSe]:EVM:AVERage[:STATe]?
Example
:EVM:AVER OFF
:EVM:AVER?
Couplings As Radio Std Detection is set to Auto, Average State is coupled to OFF
Preset
ON
State Saved Saved in instrument state
Range On|Off
Avg Mode
Selects the type of termination control used for the averaging function. This
determines the averaging action after the specified number of data acquisitions
(average count) is reached.
Exponential: After the average count is reached, each successive data
acquisition is exponentially weighted and combined with the existing average
Repeat: After reaching the average count, the averaging is reset, and a new
average is started
The default value is Exp.
Remote Command
[:SENSe]:EVM:AVERage:TCONtrol EXPonential | REPeat
[:SENSe]:EVM:AVERage:TCONtrol?
Example
:EVM:AVER:TCON REP
:EVM:AVER:TCON?
Preset
EXPonential
State Saved Saved in instrument state
Range Exponential|Repeat
969 WLAN Mode User's &Programmer's Reference

Table of Contents

Related product manuals