3 WLAN Mode
3.8 Power vs Time Measurement
2-Rise for Transmit On/Off Power
:DISP:PVT:WIND3:TRAC:X:PDIV 1ms
3-Fall for Transmit On/Off Power
Couplings If X Auto Scaling is On, this value is automatically determined by the measurement result. When you set
a value manually, X Auto Scaling automatically changes to Off.
Preset 4 ms
State Saved Saved in instrument state.
Min 1.00 ns
Max 1.0 s
Backwards Compatibility SCPI
Window Numbers used to be a combination of View and Window, now the VIEW
parameter is not used and defaults to 1. All Windows are referenced through
VIEW1. For backwards compatibility the old View and Window numbers are
honored as below:
Window Old SCPI Command New SCPI Command
PVT - Burst :DISP:PVT:VIEW[1]:WIND:TRAC:X:PDIV 1ms :DISP:PVT:WIND[1]:TRAC:X:PDIV
1ms
PVT – Rise :DISP:PVT:VIEW2:WIND:TRAC:X:PDIV 1ms :DISP:PVT:WIND2:TRAC:X:PDIV 1ms
PVT – Fall :DISP:PVT:VIEW2:WIND2:TRAC:X:PDIV 1ms :DISP:PVT:WIND3:TRAC:X:PDIV 1ms
Ref Position
Sets the reference position for the X axis to Left, Center or Right.
Remote
Command
:DISPlay:PVTime:WINDow[1]|2|3:TRACe:X[:SCALe]:RPOSition LEFT | CENTer | RIGHt
:DISPlay:PVTime:WINDow[1]|2|3:TRACe:X[:SCALe]:RPOSition?
Window numbers are as follows:
Burst: 1
Rise: 2
Fall: 3
Example
:DISP:PVT:WIND:TRACE:X:RPOS LEFT
1-Burst for Transmit On/Off Power
:DISP:PVT:WIND:TRACE:X:RPOS?
1-Burst for Transmit On/Off Power
:DISP:PVT:WIND2:TRAC:X:RPOS LEFT
2-Rise for Transmit On/Off Power
1405 WLAN Mode User's &Programmer's Reference