3 WLAN Mode
3.8 Power vs Time Measurement
:DISP:PVT:WIND:TRAC:Y:PDIV?
1-Burst for Transmit On/Off Power
:DISP:PVT:WIND2:TRAC:Y:PDIV 10
2-Rise for Transmit On/Off Power
:DISP:PVT:WIND3:TRAC:Y:PDIV 10
3-Fall for Transmit On/Off Power
Dependencies In measurements that support both Log and Lin scales, Scale/Div is grayed out in linear Y scale.
Sending the equivalent SCPI command does change the Scale/Div, though it has no affect while in
Lin.
Couplings In measurements that support Auto Scaling, when the Auto Scaling is On, this value is automatically
determined by the measurement result. When you set a value manually, Auto Scaling automatically
changes to Off.
Preset 10.00 dB / Div
State Saved Saved in instrument state
Min 0.10 dB
Max 20dB
Annotation <value> dB/ left upper of graph ! all measurements unless noted
In log scale, the Scale/Div is shown in the upper left side of the display. In Lin mode, no annotation is
displayed. ! Swept SA
Backwards Compatibility SCPI
Window Numbers used to be a combination of View and Window, now the VIEW
parameter is not used and defaults to 1. All Windows are referenced through
VIEW1. For backwards compatibility the old View and Window numbers are
honored as below:
Window Old SCPI Command New SCPI Command
Burst :DISP:PVT:VIEW[1]:WIND:TRAC:Y:PDIV 10 :DISP:PVT:WIND1:TRAC:Y:PDIV 10
Rise :DISP:PVT:VIEW2:WIND:TRAC:Y:PDIV 10 :DISP:PVT:WIND2:TRAC:Y:PDIV 10
Fall :DISP:PVT:VIEW2:WIND2:TRAC:Y:PDIV 10 :DISP:PVT:WIND3:TRAC:Y:PDIV 10
Ref Position
Positions the reference level at the top, center, or bottom of the Y Scale display.
Changing the reference position does not change the reference level value.
Remote
Command
:DISPlay:PVTime:WINDow[1]|2|3:TRACe:Y[:SCALe]:RPOSition TOP | CENTer | BOTTom
:DISPlay:PVTime:WINDow[1]|2|3:TRACe:Y[:SCALe]:RPOSition?
Window numbers are as follows: