3 WLAN Mode
3.8 Power vs Time Measurement
State Saved Yes
Min -120.00 dB
Max 0.00 dB
Ramp Down End Level
It specifies the relative value between ramp down end level and max power level
measured in ramp time length.
Remote
Command
[:SENSe]:PVTime:THReshold:DOWN:STOP <rel_ampl>
[:SENSe]:PVTime:THReshold:DOWN:STOP?
Example
:PVT:THR:DOWN:STOP -50.0
:PVT:THR:DOWN:STOP?
Preset -20.00 dB
State Saved Yes
Min -120.00 dB
Max 0.00 dB
3.8.8 Sweep
The Sweep key contains controls which allow you to control the sweep and
measurement functions of the analyzer, such as the sweep or measurement time
and whether in Single sweep/measure or Continuous sweep/measure mode.
3.8.8.1 Sweep/Control
Accesses controls that let you operate the sweep and control functions of the
instrument, such as Sweep Time and Continuous/Single.
Sweep/Measure
Lets you toggle between Continuous and Single sweep or measurement operation.
The single/continuous state is Meas Global, so the setting affects all measurements.
The front-panel key Single/Cont performs exactly the same function
See "More Information" on page 1399
Remote
Command
:INITiate:CONTinuous OFF | ON | 0 | 1
:INITiate:CONTinuous?
Example Put instrument into Single measurement operation:
WLAN Mode User's &Programmer's Reference 1398