3 WLAN Mode
3.8 Power vs Time Measurement
Example
:PVT:AVER:COUN 1
:PVT:AVER:COUN?
:PVT:AVER OFF
:PVT:AVER?
Preset 10
ON
State Saved Yes
Min 1
Max 10000
Average Type
Specifies the type of trace and result averaging to use.
This parameter is valid only for Measure Trace.
Pwr Avg (RMS) - True power averaging that is equivalent to taking the RMS
value of the voltage. It is the most accurate type of averaging
Log-Pwr Avg (Video) - Simulates the traditional spectrum analyzer type of
averaging by averaging the log of the power
Remote Command
[:SENSe]:PVTime:AVERage:TYPE LOG | RMS
[:SENSe]:PVTime:AVERage:TYPE?
Example
:PVT:AVER:TYPE LOG
:PVT:AVER:TYPE?
Preset
RMS
State Saved Yes
Range Pwr Avg (RMS)|Log-Pwr Avg(Video)
Averaging Mode
Selects the type of termination control used for the averaging function. This
determines the averaging action after the specified number of data acquisitions
(average count) is reached.
Exponential - After the average count is reached, each successive data
acquisition is exponentially weighted and combined with the existing average.
Repeat - After reaching the average count, the averaging is reset and a new
average is started.
The default value is Exponential.
1387 WLAN Mode User's &Programmer's Reference