3 WLAN Mode
3.9 Spectral Flatness Measurement
control and the INITiate:IMM command initiate a sweep/ measurement/ average sequence/hold
sequence including MaxHold and MinHold.
For Spectrum Analysis mode in ESA and PSA, the Single hardkey restarted the sweep regardless of
whether or not you were in an active sweep or sweep sequence. In the X-Series, Restart does this.
INIT[:IMM] in ESA & PSA Spectrum Analysis Mode does an implied ABORt. In some other PSA Modes,
INIT[:IMM] is ignored if not in the idle state. This conforms to SCPI specified operation for INIT[:IMM] as
shown in SCPI Command Reference section 24.7.2. The X-Series follows the ESA/PSA SA Mode
model, which may cause some Modes to have compatibility problems.
More Information
With Avg/Hold Num (in the Meas Setup menu) set to Off or set to On with a value of
1, a sweep is taken after the trigger condition is met; and the analyzer continues to
take new sweeps after the current sweep has completed and the trigger condition is
again met. However, with Avg/Hold Num set to On with a value >1, multiple sweeps
(data acquisitions) are taken for the measurement. The trigger condition must be
met prior to each sweep. The sweep is not stopped when the average count k equals
the number N set for Avg/Hold Num is reached, but the number k stops
incrementing. A measurement average usually applies to all traces, marker results,
and numeric results. But sometimes it only applies to the numeric results.
If the analyzer is in Single measurement, pressing the Cont/Single toggle control
does not change k and does not cause the sweep to be reset; the only action is to
put the analyzer into Continuous measurement operation.
If it is already in continuous sweep:
the INIT:CONT 1 command has no effect
the INIT:CONT 0 command will place the analyzer in Single Sweep but will have
no effect on the current sequence until k = N, at which point the current
sequence will stop and the instrument will go to the idle state.
If you are already in single sweep, the INIT:CONT OFF command has no effect.
If you are already in Single Sweep, then pressing the Cont/Single toggle control in
the middle of a sweep does not restart the sweep or sequence. Similarly, pressing
the Cont/Single toggle control does not restart the sweep or sequence if the sweep
is not in the idle state (for example, if you are taking a very slow sweep, or the
analyzer is waiting for a trigger). Even though pressing the Cont/Single toggle
control in the middle of a sweep does not restart the sweep, sending INIT:IMMediate
does reset it.
If in Single (and not Averaging) and you want to take one more sweep, press Restart.
If in Single and Averaging and you want to take one more sweep without resetting
the Average trace or count, simply go to Meas Setup, and increment the average
count by 1 by pressing the step up key while Average/Hold Number is the active
WLAN Mode User's &Programmer's Reference 1530