4. MEASUREMENT FLOW
3 Move the stylus to the measurement starting position, and press .
Measurement starts.
4 To prevent the damage of detector and stylus, release the stylus from a workpiece
before tilt adjustment.
<Height/tilt adjustment unit>
Raise the drive/detector unit by turning the up/down knob clockwise.
5 Adjust the level by turning the tilt knob according to the corrected value displayed in
the screen.
<Height/tilt adjustment unit>
Height/tilt adjustment unit
Tilt angle of the
drive/detector unit: ±1.5°
Tilt knob
Turning the tilt knob will change the tilt
angle of the drive/detector unit.
(Approx. 0.27° per full turn of the knob)
Screen display after DAT measurement Direction to turn the up/down knob (viewed from top)
Counterclockwise
(The left side of the drive/detector
unit in the figure is raised
comparing with the state before
DAT measurement.)
Clockwise
(The left side of the drive/detector
unit in the figure is lowered
comparing with the state before
DAT measurement.)
6 After adjustment is complete, set the stylus in place on the workpiece, then bring its
position to near the center by vertical positioning.
7 Press to perform the DAT measurement again to confirm that the leveling is
complete.
No. 99MBB765A
4-5