No. 99MBB765A
6-4
NOTE
When performing leveling with a step specimen, use a flat surface, not a step
3 Make sure that the direction of the movement of the detector is perpendicular to the
cutter mark on the specimen.
Setting positions of the drive unit and precision roughness specimen
4 Confirm that the detector is parallel to the measured surface.
Front view of detector
If the detector is not parallel to the measured surface, loosen the screw (M4) securing the
drive unit to the height/tilt adjustment unit. Adjust the detector so that it is parallel to the
measured surface, then secure the drive unit again.
Side view of detector (skidless)
If the detector is not parallel to the measured surface, turn the height/tilt adjustment on the
height-tilt adjustment unit to adjust the detector.
Side view of detector (skid-attached)
If the detector is not parallel to the measured surface, turn the height/tilt adjustment on the
height-tilt adjustment unit to adjust the detector.
Precision roughness
specimen
Overhead
view
Detector traversing
direction
Detector
Measured
surface
90°
Detector
Measured surface
Detector
Measured surface