No. 99MBB765A
6.3 Calibrating the SJ-410
If calibrating the SJ-410 with the roughness specimen supplied, be sure to calibrate with
the default values (factory settings).
■ Default value of the roughness calibration
conditions
Calibrating
condition setup
Default value
Calibration method Roughness
calibration
Number of
measurements per
charge
1 time
Nominal value 2.950 μm
(39.370 μin)
Roughness
standard
JIS1994
Filters GAUSS
λc 2.5 mm
(0.098 in)
λs NONE
Number of sampling
lengths
5
Traversal speed 1.0 mm/s
(0.039 in/s)
Returning speed 2.0 mm/s
(0.078 in/s)
Measurement range AUTO
Auto-start OFF
Auto-return ON
Retraction OFF
■ Default value
of the step calibration conditions
Calibrating
Default value
Calibration method Step calibration
Nominal value 10.000 μm
(393.700 μin)
Traversal speed 0.5 mm/s (0.019 in/s)
Returning speed 2.0 mm/s (0.078 in/s)
Measurement range 80 μm (3149.606 μin)
Auto-start OFF
Auto-return ON
Retraction OFF
NOTE
• If the SJ-410 needs to be calibrated with a standard other than the supplied precision
roughness specimen, the default calibration conditions must be modified for the
roughness specimen to be used. For information about changing the calibration
condition settings, refer to "6.4 Setting Calibration Conditions" (page 6-11).