6. CALIBRATION
No. 99MBB765A
6-9
■ Operating procedure (Refer to 6.2 "■ Accessing the Calibration Menu screen".)
Home screen to Main Menu
Calibration Measurement screen
1
Check the calibration conditions on the Calibration Measurement
screen.
If the conditions are different from the specimen, modify the
calibration conditions.
If the calibration conditions do not require modification, proceed to
the next step.
Calibration Menu screen
2
Touch on the Calibration Menu screen.
Nominal Value Setup screen
3
Set the nominal value.
TIP
Touching sets the value to 0.
For information about numeric value entry, refer to
"2.2.4 Entering numeric values/characters" (page 2-5).
4
Touch to accept the "Nom Val.".
TIP
To cancel the setting, touch .
Calibration Measurement screen
5
Touch [Close].
The set values will appear on the Calibration Measurement
screen.