6. CALIBRATION
No. 99MBB765A
■ Operating procedure (Refer to 6.2 "■ Accessing the Calibration Menu screen".)
Home screen to Main Menu ⇒
Calibration Measurement screen
Check the calibration conditions on the Calibration Measurement
screen.
If the conditions are different from the specimen, modify the
calibration conditions.
If the calibration conditions do not require modification, proceed to
the next step.
Calibration Menu screen
Touch on the Calibration Menu screen.
Nominal Value Setup screen
Set the nominal value.
TIP
• Touching sets the value to 0.
•
For information about numeric value entry, refer to
"2.2.4 Entering numeric values/characters" (page 2-5).
Touch to accept the "Nom Val.".
TIP
• To cancel the setting, touch .
Calibration Measurement screen
Touch [Close].
The set values will appear on the Calibration Measurement
screen.