No. 99MBB765A
Measurement Waveform Display screen
6 Press
.
Start calibration measurement.
A calibration measurement with the specimen is performed,
and Measurement Wav
eform Display screen is displayed
during the calibration measurement (while the detector is
traversing).
When the calibration measurement has been completed, the
measured value will be displayed in the lower column.
Calibration Measurement screen
To confirm the waveform, touch
display the waveform graph.
Measurement Waveform Display screen
The waveform graph of the calibration measurement is
displayed.
TIP
•
For information about enlarging/reducing or scrolling
waveform, refer to "4.6.2.Di
profiles" (page 4-18).
Calibration Measurement screen
Touch [Update] to update the calibration value.
TIP
• To cancel the measurement value, touch
[Cancel].
Calibration Measurement screen
The calibration factor is updated.