Remote control commands
R&S
®
RTP
1218User Manual 1337.9952.02 ─ 12
22.8.7.5 Modular probes
The commands of this chapter are relevant for R&S RT-ZM modular probes. The com-
mands are available in firmware version 3.40 and higher.
The suffix <m> selects the input channel to which the probe is connected.
TRPRobe:SETup:PRMode............................................................................................1218
PROBe<m>:SETup:PRMode.........................................................................................1218
PROBe<m>:SETup:DMOFfset...................................................................................... 1218
PROBe<m>:SETup:CMOFfset...................................................................................... 1219
PROBe<m>:SETup:NOFFset........................................................................................ 1219
PROBe<m>:SETup:POFFset........................................................................................ 1219
PROBe<m>:SETup:TERM:STATe..................................................................................1220
PROBe<m>:SETup:TERM:MODE................................................................................. 1220
PROBe<m>:SETup:TERM:MEASure?........................................................................... 1220
PROBe<m>:SETup:TERM:ADJust.................................................................................1221
TRPRobe:SETup:PRMode <PrbMeasMd>
PROBe<m>:SETup:PRMode <MeasMode>
Sets the measurement mode of the modular probe.
Suffix:
<m>
.
1..4
Parameters:
<MeasMode> DMODe | CMODe | PMODe | NMODe
DMODe
Differential mode input voltage (V
dm
), the voltage between the
positive and negative input terminal.
CMODe
Common mode input voltage (V
cm
) , the mean voltage between
the positive and negative input terminal vs. ground.
PMODe
Positive single-ended input voltage (V
p
). the voltage between the
positive input terminal and ground.
NMODe
Negative single-ended input voltage (V
N
). the voltage between
the negative input terminal and ground.
*RST:
DMODe
Usage: Asynchronous command
PROBe<m>:SETup:DMOFfset <DMOffset>
Sets the differential offset to compensate a DC voltage applied to the positive and the
negative input terminal.
Suffix:
<m>
.
1..4
Acquisition and setup