Remote control commands
R&S
®
RTP
1690User Manual 1337.9952.02 ─ 12
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set with TRIGger<m>:MILStd:CDST:RMIN.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:MILStd:CDST:RMIN and
TRIGger<m>:MILStd:CDST:RMAX.
*RST:  EQUal
TRIGger<m>:MILStd:DATA:RMIN 
<RTAPatternMin>
TRIGger<m>:MILStd:CMD:RMIN <RTAPatternMin>
TRIGger<m>:MILStd:CDST:RMIN <RTAPatternMin>
Specify a remote terminal address or set the the start value of a remote terminal
address range:
●
DATA: for data words
●
CMD: for command words
●
CDST: for status words
Parameters:
<RTAPatternMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:DATA:RMAX <RTAPatternMax>
TRIGger<m>:MILStd:CMD:RMAX <RTAPatternMax>
TRIGger<m>:MILStd:CDST:RMAX <RTAPatternMax>
Set the end value of a data range if TRIGger<m>:MILStd:CDST:RCONdition is set
to INRange or OORange:
●
DATA: for data words
●
CMD: for command words
●
CDST: for status words
Parameters:
<RTAPatternMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:CMD:CCONdition <WordCntOperator>
Sets the operator to set a specific data word count or mode code pattern.
Parameters:
<WordCntOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
Protocols