TRIGger<m>:DPHY:DATA:MAX <DataMax>
Sets the the end value of a data type range if TRIGger<m>:DPHY:DATA:CONDition
is set to INRange or OORange.
Parameters:
<DataMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:DPHY:DATA:MIN <DataMin>
Specifies a data bit pattern, or sets the the start value of a pattern range.
Parameters:
<DataMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:DPHY:DIDX:CONDition <Format>
Sets the condition for the data index. You can define an exact value or a value range.
Parameters:
<Format> EQUal | LTHan | LETHan | GTHan | GETHan | INRange |
RANGe
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set with the corresponding TRIGger<m>:DPHY:DIDX:MIN
command.
INRange = RANGe
In range: Set the minimum and maximum value using
TRIGger<m>:DPHY:DIDX:MIN and TRIGger<m>:DPHY:
DIDX:MAX.
*RST:
INRange
TRIGger<m>:DPHY:DIDX:MAX <DXSymbol>
Sets the the end value of a data index range if TRIGger<m>:DPHY:DIDX:
CONDition is set to INRange or RANge.
Parameters:
<DXSymbol> Range: 0 to 65535
Increment: 1
*RST: 0