Remote control commands
R&S
®
RTP
1851User Manual 1337.9952.02 ─ 12
TRIGger<m>:DPHY:DIDX:MIN <SymbolType>
Specifies a data index minimum, or sets the the start value of a range.
Parameters:
<SymbolType> Range: 0 to 65535
Increment: 1
*RST: 0
TRIGger<m>:DPHY:DTYPe:CONDition <Format>
Set the condition for the data type. You can define an exact value or a value range
Parameters:
<Format> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set with the corresponding TRIGger<m>:DPHY:DTYPe:MIN
command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:DPHY:DTYPe:MIN and
TRIGger<m>:DPHY:DTYPe:MAX.
*RST:
EQUal
TRIGger<m>:DPHY:DTYPe:MAX <HSDataTypeMax>
Sets the the end value of a data type range if TRIGger<m>:DPHY:DTYPe:
CONDition is set to INRange or OORange.
Parameters:
<HSDataTypeMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:DPHY:DTYPe:MIN <HSDataTypeMin>
Specifies a data type pattern, or sets the the start value of a pattern range.
Parameters:
<HSDataTypeMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
Protocols