Remote control commands
R&S
®
RTP
2031User Manual 1337.9952.02 ─ 12
SEARch:TRIGger:USB:PCONdition <SearchName>,<OperatorPort>
SEARch:TRIGger:USB:PCONdition? <SearchName>
Sets the operator to set a specific port number or a port number range. The port num-
ber values are set with SEARch:TRIGger:USB:PMIN and SEARch:TRIGger:USB:
PMAX.
Parameters:
<OperatorPort> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one port number
to be set using SEARch:TRIGger:USB:PMIN.
INRange | OORange
In range, out of range. These conditions require a port number
range to be set using SEARch:TRIGger:USB:PMIN and
SEARch:TRIGger:USB:PMAX.
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:PMIN <SearchName>,<Port>
SEARch:TRIGger:USB:PMIN? <SearchName>
Specifies a port number, or sets the start value of a port number range.
Parameters:
<Port>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:PMAX <SearchName>,<PortTo>
SEARch:TRIGger:USB:PMAX? <SearchName>
Sets the the end value of a port number range if TRIGger<m>:USB:PCONdition is
set to INRange or OORange.
Parameters:
<PortTo>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:PIDerror <SearchName>,<PIDError>
SEARch:TRIGger:USB:PIDerror? <SearchName>