Remote control commands
R&S
®
RTP
2032User Manual 1337.9952.02 ─ 12
Parameters:
<PIDError> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:SCONdition <SearchName>,<OperatorSEU>
SEARch:TRIGger:USB:SCONdition? <SearchName>
Sets the operator to set a specific SEU or an SEU range. The SEU values are set with
SEARch:TRIGger:USB:SMIN and SEARch:TRIGger:USB:SMAX.
For SEU, see "SEU" on page 818.
Parameters:
<OperatorSEU> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one SEU value
to be set using SEARch:TRIGger:USB:SMIN.
INRange | OORange
In range, out of range. These conditions require a range of SEU
values to be set using SEARch:TRIGger:USB:SMIN and
SEARch:TRIGger:USB:SMAX.
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:SMIN <SearchName>,<SEU>
SEARch:TRIGger:USB:SMIN? <SearchName>
Specifies an SEU, or sets the start value of an SEU range.
Parameters:
<SEU>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:SMAX <SearchName>,<SEUTo>
SEARch:TRIGger:USB:SMAX? <SearchName>
Sets the the end value of an SEU range if TRIGger<m>:USB:SCONdition is set to
INRange or OORange.
Parameters:
<SEUTo>
Protocols