Configuring the Internal Baseband Source
R&S
®
SMBV100B
136User Manual 1178.4460.02 ─ 03
a) Select "Baseband > Custom Digital Modulation > Marker > Marker 2 Mode >
CList".
b) Select "Baseband > Custom Digital Modulation > Marker > Global Connector
Setting".
Select "User 2 > Direction = Output", "User 2 > Signal = Baseband Marker 2".
2. To enable the R&S SMBV100B to use the Burst Gate and Level Attenuation control
signals as defined in a control list and output one of them at the User3 connector:
a) Select "Baseband > Custom Digital Modulation > Power Ramp Control >
Source > Internal"
b) Select "Baseband > Custom Digital Modulation > Marker > Global Connector
Setting".
Select "User 3 > Direction = Output", "User 3 > Signal = Lev Att A".
3.5.4.2 How to Create and Assign a Data List
The R&S SMBV100B provides the following ways to create a data list file:
To use the dedicated Data List Editor and create a file with extension *.dm_iqd,
see "To create data lists manually" on page 136
To use the tag-oriented format and create a data list file, see "To create a data list
file using tag file format" on page 175.
To use SCPI commands and create a file in binary format, see "To create a data list
in binary format" on page 176.
To create data lists manually
Use the intuitive build in Data List Editor dialog:
1. To access the "Data List Editor":
a) Select "Baseband > Custom Digital Mod > Data Source > Data List"
b) Select "Select Data List"
c) Navigate to the required directory, for example /var/user/.
Select "New" and enter "File Name" = dl.
The "Data List Editor" opens; the data list is empty.
2. Enter a sequence of 0 and 1, for example 01110101.
Generating Custom Digital Modulated Signals

Table of Contents

Related product manuals