Remote Control Commands
R&S
®
SMBV100B
807User Manual 1178.4460.02 ─ 03
Example:
SOURce1:IQ:SOURce BASeband
SOURce1:BB:IMPairment:OPTimization:MODE FAST
Manual operation: See "Optimization Mode" on page 281
12.16.4.7 SOURce:BB:GRAPhics Subsystem
This subsystem contains the commands used to setup the graphical display.
Required options
See Chapter 7.1.1, "Required Options", on page 407.
[:SOURce<hw>]:BB:GRAPhics:MODE............................................................................. 807
[:SOURce<hw>]:BB:GRAPhics:SRATe:MODE.................................................................. 807
[:SOURce<hw>]:BB:GRAPhics:SRATe:USER................................................................... 808
[:SOURce]:BB:GRAPhics:CLOSe.................................................................................... 808
[:SOURce]:BB:GRAPhics:ADD........................................................................................808
[:SOURce]:BB:GRAPhics:SOURce..................................................................................808
[:SOURce]:BB:GRAPhics:FFTLen................................................................................... 809
[:SOURce]:BB:GRAPhics:FFTFscale............................................................................... 809
[:SOURce<hw>]:BB:GRAPhics:TRIGger:SOURce.............................................................809
[:SOURce<hw>]:BB:GRAPhics:MODE <Mode>
Selects the graphics mode of the graphical signal display.
Parameters:
<Mode> IQ | VECTor | CCDF | PSPectrum | CONStellation | EYEI | EYEQ
*RST: IQ
Example:
BB:GRAP:MODE VECT
Selects the vector diagram graphics mode.
Manual operation: See "Mode" on page 416
[:SOURce<hw>]:BB:GRAPhics:SRATe:MODE <Mode>
Sets how the time resolution of the signal is determined. Maximum resolution corre-
sponds to a diagram covering the entire signal bandwidth. The higher the resolution is,
the shorter the length of the displayed signal segment will be for the specified record-
ing depth.
Parameters:
<Mode> AUTO | FULL | USER
*RST: AUTO
Example:
BB:GRAP:SRAT:MODE FULL
Sets the sample rate mode.
Manual operation: See "Sample Rate Mode" on page 417
SOURce Subsystem