Configuring the RF Signal
R&S
®
SMBV100B
393User Manual 1178.4460.02 ─ 03
Shows all R&S NRP sensors connected to the instrument, or in the LAN.
Chapter 6.12.4.3, "NRP Power Viewer", on page 395
A function for measuring and monitoring the RF output power or a user-defined sig-
nal source.
Connecting R&S NRP Power Sensors to the R&S SMBV100B............................393
NRP Sensor Mapping........................................................................................... 394
NRP Power Viewer............................................................................................... 395
6.12.4.1 Connecting R&S NRP Power Sensors to the R&S SMBV100B
R&S NRP sensors are connected to the R&S SMBV100B in the following ways:
Connection to the Sensor connector
R&S NRP-ZK6 (six-pole interface cable) for R&S NRPxx power sensors
No additional cable for R&S NRP-Zxx power sensors (cable is fixed on the sen-
sor)
Connection to the USB connector
Requires the following cables, depending on the used sensor type:
R&S NRP-ZKU (USB interface cable) for R&S NRPxx power sensors
R&S NRP-Z3 or R&S NRP-Z4 (USB adapter cables) for sensors of the R&S
NRP-Zxx family
Connection via R&S NRP-Z5 sensor hub
The R&S NRP-Z5 USB sensor hub (high-speed USB 2.0) can host up to 4 R&S
NRP sensors. It provides simultaneous internal and external triggering of all con-
nected sensors.
Requires additional cables, depending on the used output connector of the hub.
Choose one of the following:
Short extension cable R&S NRP-Z2 for connection to the sensor connector.
This six-pole connection provides the external trigger capability.
Standard USB cable (USB type A to USB type B) to any USB type A connector
of the R&S SMBV100B. This connection does not support external triggering.
Connection via USB hub with external power supply unit
Requires the following cables, depending on the used sensor type:
R&S NRP-ZKU (USB interface cable) for R&S NRPxx power sensors
R&S NRP-Z3 or R&S NRP-Z4 (USB adapter cables) for sensors of the R&S
NRP-Zxx family
Connection via LAN for R&S NRPxxxSN power sensors
Using the Ethernet interface requires PoE (Power over Ethernet) to provide the
electrical power.
To establish the connection, you can use:
A PoE Ethernet switch, e.g. R&S NRP-ZAP1 and an RJ-45 Ethernet cable.
A PoE injector and an RJ-45 Ethernet cable.
For details, see the description R&S
®
NRP
®
Series Power Sensors Getting Started.
Improving Level Performance