Remote Control Commands
R&S
®
SMBV100B
681User Manual 1178.4460.02 ─ 03
[:SOURce]:INPut:USER:CLOCk:SLOPe <Slope>
Sets the polarity of the active slope of an externally applied clock signal.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Manual operation: See "Clock Input Slope" on page 524
[:SOURce]:INPut:USER:TRIGger:SLOPe <Slope>
Sets the polarity of the active slope of an applied instrument trigger.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Example: See Example "Global connectors settings configuration"
on page 677
Manual operation: See "Trigger Input Slope" on page 524
12.16.2 SOURce:BBIN Subsystem
The SOURce:BBIN remote control commands will be supported in a future firmware
version.
The baseband input cannot be activated.
The baseband input related commands in this section and in the SCONfiguration Sub-
system have no effect because the baseband input interfaces cannot be activated.
The SOURce:BBIN subsystem contains the commands for setting the external digital
baseband signal.
Required options
See "Required Options" on page 456.
Example: Applying an external digital baseband signal (Dig I/Q)
An external digital signal must be applied at the Dig I/Q connectors.
// set the sample rate source and query the value
SOURce1:BBIN:SRATe:SOURce DIN
SOURce1:BBIN:SRATe:ACTual?
SOURce1:BBIN:STATe ON
// enable automatic adjustment of the baseband input signal
SOURce1:BBIN:DIGital:ASETting:STATe ON
SOURce Subsystem