Configuring the RF Signal
R&S
®
SMBV100B
379User Manual 1178.4460.02 ─ 03
Risk of DUT damage due to high input power
The settings combination "ALC > Sample Level > Full" and "ALC > State > Off (Sample
&Hold)" or "ALC > State > Sample & On" results in signals with large headroom. With
this settings combination, the R&S SMBV100B generates a short sample pulse with full
RF output power.
To protect the DUT from damage due to high input power, use the setting "Sample
Level > Attenuated" or "Sample Level > Minimum".
6.12.2.1 ALC Settings
Access:
Select "RF" > "RF Level" > "Automatic Level Control".
In the "ALC" dialog, you can configure the settings for the automatic level control of
the RF signal to achieve optimal accuracy.
The remote commands required to define these settings are described in Chap-
ter 12.16.15, "SOURce:POWer Subsystem", on page 867.
Settings
State............................................................................................................................379
Detector Sensitivity..................................................................................................... 380
Sample Level.............................................................................................................. 381
State
Selects the internal level control mode.
Improving Level Performance