Configuring the RF Signal
R&S
®
SMBV100B
357User Manual 1178.4460.02 ─ 03
Remote command:
[:SOURce<hw>]:AM:DEViation:MODE on page 693
Total Depth ← AM settings
Sets the sum of the signal Depth for path 1 and path 2 in per cent when using com-
bined sources and Deviation Mode > Fixed Total.
Using external LF signals, the R&S SMBV100B indicates the Sensitivity per Volts for
the input signal.
Provided both signal paths are turned on, the depth of both paths always sum up to the
value of the total depth.
Behavior when changing any depth settings:
●
Changing the total depth adjusts the depths of both paths proportionally.
●
Changing the depth of one path adjusts the depth of the other path. The sum
always results in the set total depth.
Remote command:
[:SOURce<hw>]:AM:DEPTh:SUM on page 693
Ratio Path2/Path1 ← AM settings
Sets the deviaton ratio of the signal depth for path 2 to path 1 in per cent when using
combined sources.
Using external LF signals, the R&S SMBV100B indicates the Sensitivity per Volts for
the input signal.
Remote command:
[:SOURce<hw>]:AM:RATio on page 694
6.11.4.3 Pulse Generator Settings
Option: see Chapter 6.11.1, "Required Options", on page 346.
Access:
► Select "RF" > "Analog Modulation" > "Pulse Generator".
Analog Modulations