Multi-Instrument Setups
R&S
®
SMBV100B
453User Manual 1178.4460.02 ─ 03
Remote command:
:SCONfiguration:EXTernal:DIGital:REMote:ISELect on page 674
:SCONfiguration:EXTernal:DIGital:REMote:CONNect on page 674
:SCONfiguration:EXTernal:IQOutput:REMote:ISELect on page 674
:SCONfiguration:EXTernal:IQOutput:REMote:CONNect on page 674
Remote Control Settings
Access:
1. Select "Taskbar > System Config > System Configuration > External RF and I/Q".
2. Navigate to the required connector.
Select "External Instruments > Config ...".
3. In the "<Dig. Conn. Name>: External Instrument Configuration" dialog, select
"Remote Control".
In this dialog, you can send individual SCPI commands or a sequence of SCPI
commands to the connected instruments.
The sent commands and their results are displayed in the "Command Trace" field, as
long as the trace is not deleted "Clear Trace". The sent commands are also listed in
the "History" list. As long as this history list is not deleted ("Clear History"), you can
select a command from the list and send it again.
Settings:
Mode........................................................................................................................... 454
Command, Send Command........................................................................................454
File, Sequence, Send Sequence.................................................................................454
History, Clear History...................................................................................................454
Commands Trace, Clear Trace................................................................................... 454
System Configuration Settings