Remote Control Commands
R&S
®
SMBV100B
881User Manual 1178.4460.02 ─ 03
Options: DER1G requires R&S SMBVB-B3
Manual operation: See "Reference Output/1GHz Reference Output" on page 304
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue <Value>
Specifies the frequency correction value (adjustment value).
Parameters:
<Value> integer
*RST: 0
Example: See [:SOURce]:ROSCillator[:INTernal]:ADJust[:
STATe] on page 881
Manual operation: See "Adjustment Value" on page 305
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe]
<State>
Determines whether the calibrated (off) or a user-defined (on) adjustment value is
used for fine adjustment of the frequency.
Parameters:
<State> 0 | 1 | OFF | ON
0
Fine adjustment with the calibrated frequency value
1
User-defined adjustment value.
The instrument is no longer in the calibrated state.
The calibration value is, however, not changed. The instrument
resumes the calibrated state if you send
SOURce:ROSCillator:INTernal:ADJust:STATe 0.
*RST: n.a. (factory preset: 0)
Example: see Example "Configuring the reference oscillator" on page 877.
Manual operation: See "Adjustment Active" on page 305
12.16.17 SOURce:SWEep Subsystem
The SOURce:SWEep subsystem contains the commands for configuring RF sweep sig-
nals.
SOURce Subsystem