Remote Control Commands
R&S
®
SMBV100B
662User Manual 1178.4460.02 ─ 03
Manual operation: See " Sensor type and serial number " on page 399
:SENSe<ch>[:POWer]:ZERO
Performs zeroing of the sensor.
Zeroing is required after warm-up, i.e. after connecting the sensor.
Note: Switch off or disconnect the RF power source from the sensor before zeroing.
We recommend that you zero in regular intervals (at least once a day), if:
●
The temperature has varied more than about 5 °C.
●
The sensor has been replaced.
●
You want to measure very low power.
Example:
SENS1:ZERO
Executes zeroing.
Usage: Event
Manual operation: See " Zero " on page 399
12.15 SCONfiguration Subsystem
The SCONfiguration subsystem contains the commands for defining the system
configuration setting, like stream mapping and used baseband configuration.
Required options
See "Required options" on page 442.
Example: Distributing the I/Q streams to the output connectors
The following example lists the commands necessary to fulfill this task.
// ******************************************************************
// Map the I/Q streams to the output connectors
// ******************************************************************
SCONfiguration:OUTPut:MAPPing:RF1:STReam1:STATe ON
SCONfiguration:OUTPut:MAPPing:IQOutput1:STReam3:STATe ON
Example: Connecting and configuring external instrument
The following example lists the commands necessary to fulfill this task.
SCONfiguration:EXTernal:DISPlay ALL
SCONfiguration:EXTernal:REMote:CLEan
SCONfiguration:EXTernal:REMote:SCAN
SCONfiguration:EXTernal:REMote:SCAN:STATe?
// 0
SCONfiguration Subsystem