Remote Control Commands
R&S
®
SMBV100B
725User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:BB:DM:PRAMp[:STATe] <State>
Enables or disables power ramping.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Enable power ramping" on page 714
Manual operation: See "State" on page 127
Trigger Settings
[:SOURce<hw>]:BB:DM[:TRIGger]:SEQuence <Sequence>
Selects the trigger mode. For detailed description of the trigger modes, refer to "Impact
of the Trigger Modes on the Signal Generation" on page 97.
Parameters:
<Sequence> AUTO | RETRigger | AAUTo | ARETrigger | SINGle
*RST: AUTO
Example: See Example "Adjusting clock, marker and trigger settings"
on page 713
Manual operation: See "Mode" on page 110
[:SOURce<hw>]:BB:DM:TRIGger:SOURce <Source>
Selects the trigger signal source and determines the way the triggering is executed.
Provided are:
●
Internal triggering by a command (INTernal)
●
External trigger signal via one of the User x connectors
– EGT1: External global trigger
– EGC1: External global clock
●
In master-slave mode, the external baseband synchronization signal (BBSY)
●
EXTernal: Setting only
Provided only for backward compatibility with other Rohde & Schwarz signal gener-
ators.
The R&S SMBV100B accepts this value and maps it automatically as follows:
EXTernal = EGT1
Parameters:
<Source> INTernal|EGT1|EGC1|EXTernal|BBSY
*RST: INTernal
Example: See Example "Adjusting clock, marker and trigger settings"
on page 713
Manual operation: See "Source" on page 111
SOURce Subsystem