Troubleshooting and Error Messages
R&S
®
SMBV100B
950User Manual 1178.4460.02 ─ 03
Static Errors/Error History
Toggles between "Static" and "History" view of the info dialog.
Remote command:
:SYSTem:ERRor[:NEXT]? on page 894
Each time a SYST:ERR:NEXT? query is sent, the eldest entry in the error queue is
returned and at the same time cleared in the list.
:SYSTem:ERRor:STATic? on page 895
Queries the list of all errors.
Clear History
Clears all messages in the "History" view.
Remote command:
:SYSTem:ERRor:ALL? on page 892
Each time a SYST:ERR:ALL? query is sent, the error queue is returned and at the
same time cleared.
:SYSTem:ERRor:HISTory:CLEar on page 894
Clears the messages in the "History" view.
14.5 Resolving Network Connection Failures
Several issues may cause failures in the network connection to the instrument. This
section lists the most likely reasons and the recommended solutions.
Common reasons for network connection failures
●
Network connecting cables and cable connectors of poor quality
●
Incompatibility between the network interface of the R&S SMBV100B and certain
switches or routers available on the market
●
An invalid IP address assigned to the instrument
Possible solutions
Risk of network failure
Consult your network administrator before performing the following tasks:
●
Connecting the instrument to the network
●
Configuring the network
●
Changing IP addresses
Errors can affect the entire network.
●
Check the network infrastructure. Exchange connecting cables if obvious damage
is visible.
Resolving Network Connection Failures