Configuring the RF Signal
R&S
®
SMBV100B
293User Manual 1178.4460.02 ─ 03
Settings
Amplitude.................................................................................................................... 293
Limit ............................................................................................................................293
Offset ..........................................................................................................................294
Digital Attenuation.......................................................................................................294
Setting Characteristics ............................................................................................... 294
Mode .......................................................................................................................... 295
Level Range ...............................................................................................................295
Readjust......................................................................................................................296
User Variation..............................................................................................................296
└ Variation Active............................................................................................. 296
└ Variation Step................................................................................................296
Amplitude
Sets the level of the RF signal.
The value is offset-free and corresponds to the level at the RF connector.
See "RF frequency and level display with a downstream instrument" on page 288.
Note: Suppressed values in the status bar
For security concerns or certain operating modes, you can hide the frequency and
level display in the status bar.
●
The display has been disabled for security reasons.
See:
– Annotation Frequency
– Annotation Amplitude
The display is disabled when:
– List mode is running, see Chapter 6.10, "Varying the RF Signal in List or Sweep
Mode", on page 310.
Remote command:
[:SOURce<hw>]:POWer:POWer on page 872
Note: The SCPI command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude] sets the level of the "Level" display.
This means, the level containing offset.
Limit
Sets an upper limit for the RF output power.
You can use this value to protect your DUT from damage due to high input power. If
you enter an RF level above this value, the instrument limits the output power to this
specified value, and generates the warning message:
"Pep value greater than the defined limit." However, the level indication in the status
bar is not affected.
The setting is not affected by an instrument preset ([Preset] key or *RST) and the
"Save/Recall" function. It is reset only by factory preset.