Remote Control Commands
R&S
®
SMBV100B
924User Manual 1178.4460.02 ─ 03
:TRIGger<hw>:FSWeep[:IMMediate]
:TRIGger<hw>:PSWeep[:IMMediate]
:TRIGger<hw>:LFFSweep:IMMediate
:TRIGger<hw>[:SWEep][:IMMediate]
Performs a single sweep and immediately starts the activated, corresponding sweep:
FSWeep - RF frequency
PSWeep - RF level
LFFSweep - LF frequency
SWEep - all sweeps
Effective in the following configuration:
TRIG:FSW|LFFS|PSW|[:SWE]:SOUR SING
SOUR:SWE:FREQ|POW:MODE AUTO or SOUR:LFO:SWE:[FREQ:]MODE AUTO
Alternatively, you can use the IMMediate command instead of the respective SWEep:
[FREQ:]|POW:EXECute command.
Example:
TRIG
Starts all active sweeps.
Usage: Event
Manual operation: See " Execute Single Sweep " on page 331
:TRIGger<hw>:LFFSweep
Executes an LF frequency sweep in the following configuration:
TRIG:LFFS:SOUR SING
LFO:SWE:MODE AUTO
Example:
LFO:SWE:MODE AUTO
TRIG:LFFS:SOUR SING
TRIG:LFFSweep
Usage: Event
12.21 UNIT Subsystem
The UNIT subsystem is used to set default units for parameters if no unit is indicated in
a command. These settings are valid for the entire instrument.
Example: Setting default units for remote control
UNIT:POW V
UNIT:ANGL DEG
Sets V (volts) as unit of all power parameters, DEG (degrees) for the phase modulation
angle and KMH for the speed.
UNIT Subsystem