Remote Control Commands
R&S
®
SMBV100B
649User Manual 1178.4460.02 ─ 03
Example:
OUTPut1:AMODe FIXed
OUTPut1:AFIXed:RANGe:UPPer?
// -27
OUTPut1:AFIXed:RANGe:LOW?
// -50
Usage: Query only
Manual operation: See " Level Range " on page 377
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped.
To define the output state, use the command :OUTPut<hw>[:STATe].
Example:
OUTP:PROT:CLE
Resets the protective circuit of the RF output.
Usage: Event
Manual operation: See "Overload" on page 377
:OUTPut<hw>:PROTection:TRIPped?
Queries the state of the protective circuit.
Return values:
<Tripped> 0 | 1 | OFF | ON
*RST: 0
Example:
OUTP:PROT:TRIP
queries the state of the protective circuit of the RF output.
Response: 0
the protective circuit has not tripped.
Usage: Query only
Manual operation: See "Overload" on page 377
12.14 SENSe, READ, INITiate and SLISt Subsystems
These subsystems contain the commands for configuring the power measurements
with R&S NRP power sensor connected to the R&S SMBV100B.