Remote Control Commands
R&S
®
SMBV100B
795User Manual 1178.4460.02 ─ 03
Parameters:
<Delay> float
Range: 0 to 16777215
Increment: 0.001
*RST: 0
Example: See "Programming Examples" on page 779
Manual operation: See "Marker x Delay" on page 114
Clock Settings
[:SOURce<hw>]:BB:MCCW:CLOCk:SOURce................................................................... 795
[:SOURce<hw>]:BB:MCCW:CLOCk:SOURce <Source>
Selects the clock source:
INTernal: Internal clock reference
Parameters:
<Source> INTernal
*RST: INTernal
Example: See "Programming Examples" on page 779.
Manual operation: See "Clock Source" on page 115
12.16.4.5 SOURce:AWGN Subsystem
The SOURce:AWGN subsystem contains the commands for setting the noise generator.
Required options
See Chapter 4.1.1, "Required Options", on page 255.
Programming Examples
Example: Generating a pure noise signal
The following example generates a pure noise signal with specified bandwidth and
noise level.
SOURce1:AWGN:MODE ONLY
SOURce1:AWGN:BWIDth 3840000
SOURce1:AWGN:BWIDth:RATio 2
SOURce1:AWGN:STATe ON
SOURce1:AWGN:BWIDth:NOISe?
// Response: 7680000
SOURce1:AWGN:DISP:MODE RFA
SOURce1:AWGN:POWer:NOISe -80
// SOURce1:POWer:LEVel:IMMediate:AMPlitude?