Multi-Instrument Setups
R&S
®
SMBV100B
463User Manual 1178.4460.02 ─ 03
"HS Digital I/Q In"
Estimates the sample rate value based on the information transmitted
by the transmitter.
Resulting value per channel is displayed by the parameter Sample
Rate.
Remote command:
[:SOURce<hw>]:BBIN:SRATe:SOURce on page 686
Sample Rate Value
For Dig I/Q connectors, sets the sample rate of the external digital baseband input sig-
nal.
The max. sample rate depends on the connected transmitting device.
Remote command:
[:SOURce<hw>]:BBIN:SRATe[:ACTual]? on page 686
8.6.4 Input Level Settings
Access:
1. Select "BB Input > Digital I/Q In > Baseband Input Settings > Interface = DIG I/Q ".
2. Select "Input Level".
DIG IQ Auto Setting.................................................................................................... 463
Measurement Period...................................................................................................464
Auto Level Set.............................................................................................................464
Crest Factor................................................................................................................ 464
Peak Level.................................................................................................................. 464
Level............................................................................................................................464
DIG IQ Auto Setting
Activates an automatic adjustment of the baseband input signal.
The R&S SMBV100B receives PEP, level and crest factor values directly from the con-
nected transmitter and recognizes changes automatically.
Remote command:
[:SOURce<hw>]:BBIN:DIGital:ASETting:STATe on page 686
Digital Baseband Input Settings