Configuring the RF Signal
R&S
®
SMBV100B
289User Manual 1178.4460.02 ─ 03
parameters, but considers all additional parameters concerning the frequency and
level.
Example parameters that affect the frequency and level:
Frequency offset
Multiplication factor
User correction
Digital attenuation
Any of these cases are indicated by a dedicated icon in the status bar, depending on
the affected parameter.
Figure 6-1 illustrates the calculation of the "Frequency" and "Level" values, displayed
in the status bar.
Figure 6-1: Frequency and Level display vs frequency and level at the RF output
f
RF
, Level
RF
= Frequency and level ("Status bar > Frequency and Level") at the output of the downstream
instrument, see also Calculation of f
RF
and Level
RF
)
f
RF Out
, Level
RF
Out
= Frequency and level at the output connector [RF] ("RF > RF Frequency > Frequency" and
"RF > RF Level > Amplitude")
n*f = Multiplication factor ("RF > RF Frequency > Multiplier")
Δf = Frequency offset ("RF > RF Frequency > Offset")
ΔLevel = Power offset ("RF > RF Level > Offset")
Calculation of f
RF
and Level
RF
f
RF
=n*f
RF Out
+ Δf
Level
RF
= Level
RF Out
+ ΔLevel + Level
DigAtt
(Level
DigAtt
for active internal I/Q modulation, "RF > RF Level > Digital Attenuation")
6.5 RF Frequency Settings
Access:
Select "RF" > "RF Frequency" > "Frequency".
RF Frequency Settings