Remote Control Commands
R&S
®
SMBV100B
632User Manual 1178.4460.02 ─ 03
:CALibration<hw>:IQModulator:LOCal?
Starts adjustment of the I/Q modulator for the currently set frequency and baseband
gain. The I/Q modulator is adjusted with respect to carrier leakage, I/Q imbalance and
quadrature.
Return values:
<Local> 0 | 1 | OFF | ON
*RST: 0
Example:
// Start adjustment for the I/Q modulator
// for the currently set frequency
CALibration:IQModulator:LOCal?
// 0
// Adjustments are performed successfully
Usage: Query only
Manual operation: See "Adjust I/Q Modulator Current Frequency" on page 279
12.8 DIAGnostic Subsystem
The DIAGnostic subsystem contains the commands used for instrument diagnosis
and servicing. SCPI does not define any DIAGnostic commands; the commands lis-
ted here are all device-specific. All DIAGnostic commands are query commands
which are not affected by *RST.
The test functions are intended for services purposes.
They are thus password-protected functions. Unlock the corresponding protection level
to access them, see :SYSTem:PROTect<ch>[:STATe]
For more information, see R&S SMBV100B Service Manual.
Common suffixes
The following common suffixes are used in remote commands:
Suffix Value range Description
DIAGnostic<hw>
[1] Optional suffix
DIAGnostic Subsystem