Configuring the RF Signal
R&S
®
SMBV100B
303User Manual 1178.4460.02 ─ 03
Remote command:
[:SOURce]:ROSCillator:EXTernal:FREQuency on page 878
Variable Reference Frequency
Option: R&S SMBVB-K704
Sets the variably settable external reference frequency.
Remote command:
[:SOURce]:ROSCillator:EXTernal:FREQuency:VARiable on page 878
Synchronization Bandwidth
("External Reference Frequency = 10 MHz, 100 MHz or 1 GHz")
Selects the synchronization bandwidth for an external reference signal.
The resulting bandwidth is indicated with the parameter Nominal Synchronization
Bandwidth.
"Narrow"
The internal reference oscillator is synchronized to the external signal
with narrow bandwidth (< 1 Hz). This setting is recommended if the
phase noise of the external signal is worse than the phase noise of
the internal OCXO.
"Wide"
This mode is the recommended standard mode and for precise refer-
ence sources of high spectral purity.
Remote command:
[:SOURce]:ROSCillator:EXTernal:SBANdwidth on page 879
Nominal Synchronization Bandwidth
Indicates the nominal synchronization bandwidth for the selected external reference
frequency and synchronization bandwidth.
Remote command:
[:SOURce]:ROSCillator:EXTernal:NSBandwidth? on page 879
Minimum Locking Range
Indicates the minimum locking range for the selected external reference frequency.
Remote command:
[:SOURce]:ROSCillator:EXTernal:MLRange? on page 879