Remote Control Commands
R&S
®
SMBV100B
878User Manual 1178.4460.02 ─ 03
[:SOURce]:ROSCillator:PRESet
Resets the reference oscillator settings.
Example: See Example "Configuring the reference oscillator" on page 877.
Usage: Event
Manual operation: See "Set to Default" on page 302
[:SOURce]:ROSCillator:SOURce <Source>
Selects between internal or external reference frequency.
Parameters:
<Source> INTernal | EXTernal
*RST: n.a. (factory preset: INTernal)
Example: See Example "Configuring the reference oscillator" on page 877.
Manual operation: See "Source" on page 302
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe] <State>
Determines that the RF output is turned off when the external reference signal is
selected, but missing.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example: See Example "Configuring the reference oscillator" on page 877.
Manual operation: See "Deactivate RF Output (if external reference is missing)"
on page 302
[:SOURce]:ROSCillator:EXTernal:FREQuency <Frequency>
Sets the frequency of the external reference.
Parameters:
<Frequency> 100MHZ | 1GHZ | VARiable | 10MHZ
*RST: n.a. (factory preset: 10MHZ)
Example: See Example "Configuring the reference oscillator" on page 877.
Options: VARiable requires R&S SMBVB-K704
1GHZ requires R&S SMBVB-B3
Manual operation: See "External Reference Frequency" on page 302
[:SOURce]:ROSCillator:EXTernal:FREQuency:VARiable <Frequency>
Specifies the user-defined external reference frequency.
SOURce Subsystem