Configuring the Internal Baseband Source
R&S
®
SMBV100B
207User Manual 1178.4460.02 ─ 03
"Disable"
No additional marker is generated.
"Marker 1, 2, 3"
Generates a restart marker signal at the beginning of the first seg-
ment of the complete multi-segment sequence.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:MARKer:FSEGment
on page 757
Segment Restart
Enables/disables the generation of an extra restart marker in the output waveform file.
If additional marker generation is enabled, the existing marker signals in the individual
segment waveform files are not considered, see also Chapter 3.7.2.4, "Impact of the
Marker Settings", on page 195.
"Disable"
No additional marker is generated.
"Marker 1, 2, 3"
Generates a restart marker signal at the beginning of each segment.
The segment begin is defined by the low-high slope of the marker sig-
nal, which applies for switching between two segments and if there is
a segment replay.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:MARKer:ESEGment
on page 757
3.7.3.4 Multi-Segment Waveform Sequencing
Access:
1. Select "Baseband > ARB > General > Multi Segment".
2. In the "ARB: Multi Segment > General" dialog, select an existing list.
3. Select "ARB: Multi Segment > Segment Table" and configure a waveform with
more than one segment.
4. Select "ARB: Multi Segment > Level/Marker/Clock" and adjust the clock settings so
that all segments use the same clock rate.
5. Select "ARB: Multi Segment > General" > "Create Output File/Create and Load
Output File".
6. Select "Sequencing List..."
7. Select "New" or "Load" to load a sequencing list file.
Generating Multi-Segment Waveform Files