EasyManua.ls Logo

Rohde & Schwarz SMBV100B - References

Rohde & Schwarz SMBV100B
1030 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring the Internal Baseband Source
R&S
®
SMBV100B
137User Manual 1178.4460.02 ─ 03
Figure 3-11: Example of data list
3. Select "Save" to store the used settings as a data list file.
The instrument stores the dl.dm_iqd file in the /var/user/ directory.
How to assign and activate data lists
Irrespectively on the way they are created, generated data lists are not automatically
used.
1. To enable the R&S SMBV100B to use the data list as data source for the custom
digital modulation:
a) Select "Baseband > Custom Digital Modulation > Data Source > Data List"
b) Select "Select Data List"
c) Navigate through the file system. Select the file and confirm with "Select".
2. To enable the R&S SMBV100B to use the data list as data source for any of the
digital standards:
a) Select the "Data List Name" in the individual dialog.
For example, for "Baseband > 3GPP FDD > Basestations > BS1 > Channel
Table > P-CCPCH > Data = Data List"
b) Select "DList = None".
c) Select the file in the file system.
3.5.5 References
Table 3-5: Custom digital modulation default values
Parameter Value
State Not affected
Data Source PRBS 9
Generating Custom Digital Modulated Signals

Table of Contents

Related product manuals