General Instrument Functions
R&S
®
SMBV100B
539User Manual 1178.4460.02 ─ 03
Secure Update Policy
Allows you to configure the automatic signature verification for firmware installation.
To apply the change: enter the security password and confirm with "Accept". Otherwise
the change has no effect.
See also:
Chapter 10.5, "Using the Security Settings", on page 535 for more information on
the security concept.
The release notes for details on signature verification when installing new or former
firmware versions, available at www.rohde-schwarz.com/firmware/smbv100b.
"Confirm Unsigned"
Performs the signature verification.
If the check detects any discrepancies, the instrument issues a warn-
ing message. You can still update the firmware or reject updating.
This setting also enables you to downgrade the firmware version.
"All Packages"
Accepts all packages without signature verification.
"R&S Signed Packages"
Performs the signature check.
If the check detects any discrepancies, the instrument issues a warn-
ing message and locks the update to this firmware.
Remote command:
:SYSTem:SECurity:SUPolicy on page 905
10.5.2.2 Disk & Memory Security Settings
Access:
Select "System Config > Setup > Security > Security > General > Disk & Memory".
The "Disk & Memory" tab secures controlled access to the mass memory and pre-
vents information from leaving a secure environment.
The remote commands available to control security settings are described in Chap-