Remote Control Commands
R&S
®
SMBV100B
629User Manual 1178.4460.02 ─ 03
Return values:
<TotalPhysMemKb> integer
Total physical memory.
<ApplicMemKb> integer
Application memory.
<HeapUsedKb> integer
Used heap memory.
<HeapAvailableKb> integer
Available heap memory.
Usage: Query only
12.7
CALibration Subsystem
The CALibration subsystem contains the commands needed for performing internal
adjustments. This procedure is triggered by the query commands.
Suffix <hw>
Suffix Value range Description
CALibration<hw>
[1] Optional suffix
Understanding the query response
0: error-free execution of the adjustments
1: indicates that an error occurred; the process has been canceled
:CALibration:ALL[:MEASure]?......................................................................................... 629
:CALibration<hw>:ALL:DATE?.........................................................................................630
:CALibration<hw>:ALL:INFormation?............................................................................... 630
:CALibration<hw>:ALL:TEMP?........................................................................................ 630
:CALibration<hw>:ALL:TIME?......................................................................................... 631
:CALibration:DATA:FACTory:DATE?.................................................................................631
:CALibration<hw>:CONTinueonerror................................................................................631
:CALibration<hw>:IQModulator:LOCal?............................................................................632
:CALibration:ALL[:MEASure]? [<Force>]
Starts all internal adjustments that do not need external measuring equipment.
Query parameters:
<Force> string
Return values:
<Measure> 0 | 1 | OFF | ON
CALibration Subsystem