Maintenance
R&S
®
SMBV100B
934User Manual 1178.4460.02 ─ 03
Contact the Rohde & Schwarz customer support, see Chapter 14.6, "Collecting
Information for Technical Support", on page 951.
13.3.3 Internal Adjustment Settings
Access:
► Select "System Config > Setup > General > Internal Adjustments".
In this dialog, you can perform internal calibration routines and get information on
the last performed calibration.
The remote commands required to define these settings are described in Chapter 12.7,
"CALibration Subsystem", on page 629.
The background color of "Temperature Offset" depends on the deviation of the temper-
ature at the beginning of the adjustment, compared to the last adjustment. We recom-
mend that you perform internal adjustments if the instrument is warmed up, but
depending on instrument usage, less than 30 minutes are sufficient.
Further functions and individually targeted calibration routines are also available but
password-protected.
●
Calibration routines
Self-calibration routines that require additional (external) equipment are performed
in the Rohde & Schwarz service center.
These functions are password-protected and require higher-level protection pass-
word, see "Protection levels" on page 536
For more information, see R&S SMBV100B Service Manual.
Adjust All
Performs all available internal calibration routines of the instrument.
Performing Maintenance Tasks