Remote Control Commands
R&S
®
SMBV100B
905User Manual 1178.4460.02 ─ 03
Example:
SYST:IDEN USER
// Selects a user-defined identification
SYST:ORES "Test Option"
// Defines the OPT string 'test option'
*OPT?
// Response: 'test option'
Manual operation: See "OPT String" on page 568
:SYSTem:LANGuage <Language>
Sets the remote control command set.
Parameters:
<Language> string
Example:
SYSTem:LANGuage "SCPI"
// selects SCPI command set
Manual operation: See "Language" on page 567
:SYSTem:INFormation:SCPI <InfoString>
Inserts system information in recorded SCPI command lists, for example information
on a missing command.
Parameters:
<InfoString> string
Example:
SYST:INF:SCPI "missing command"
enters the information into a recorded SCPI command list.
:SYSTem:SECurity:SANitize[:STATe] <SecPassWord>, <MmemProtState>
Sanitizes the internal memory.
Parameters:
<MmemProtState> 0 | 1 | OFF | ON
*RST: 0
Setting parameters:
<SecPassWord> string
Example:
SYSTem:SECurity:SANitize[:STATe] 1
Manual operation: See "Sanitize" on page 540
:SYSTem:SECurity:SUPolicy <SecPassWord>, <Update Policy>
Configures the automatic signature verification for firmware installation.