Remote Control Commands
R&S
®
SMBV100B
904User Manual 1178.4460.02 ─ 03
USER
User-defined strings can be selected.
*RST: n.a. (factory preset: AUTO)
Example:
SYST:IDEN AUTO
Automatically assigns the OPT and IDN strings according to the
selected instrument language.
Manual operation: See "Mode" on page 567
:SYSTem:IDENtification:PRESet
Sets the *IDN and *OPT strings in user defined mode to default values.
Example:
SYST:IDEN USER
SYST:IDEN:PRES
Usage: Event
Manual operation: See "Set to Default" on page 568
:SYSTem:IRESponse <IdnResponse>
Defines the user defined identification string for *IDN.
Note: While working in an emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command SYST:IRES is discarded.
Parameters:
<IdnResponse> string
Example:
SYST:IDEN USER
// Selects a user-defined identification
SYST:IRES "Test Device"
// Defines identification string 'test device'
*IDN?
// Response: 'test device'
Manual operation: See "IDN String" on page 568
:SYSTem:ORESponse <OResponse>
Defines the user defined response string for *OPT.
Note: While working in an emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command SYST:ORES is discarded.
Parameters:
<OResponse> string
SYSTem Subsystem