Remote Control Commands
R&S
®
SMBV100B
896User Manual 1178.4460.02 ─ 03
Example:
SYST:DLOC ON
Activates the display lock. The instrument cannot be operated
via the display until it has been enabled with SYST:DLOC OFF.
Manual operation: See "User Interface" on page 541
:SYSTem:KLOCk <State>
Disables the front panel keyboard of the instrument including the Local key.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
SYST:KLOC ON
Locks the front panel and external controls.
To enable the controls, send SYST:KLOC OFF.
Manual operation: See "User Interface" on page 541
:SYSTem:SAV <Pathname>
Saves the current instrument settings to a file with defined filename.
Setting parameters:
<Pathname> string
Example:
SYSTem:SAV
"/var/user/Instrument_settings.svrcltxt"
Usage: Setting only
:SYSTem:RCL <Pathname>
Loads a file with previously saved instrument settings.
Setting parameters:
<Pathname> string
Example:
SYSTem:RCL
"/var/user/Instrument_settings.svrcltxt"
Usage: Setting only
:SYSTem:PROTect<ch>[:STATe] <State>[, <Key>]
Activates and deactivates the specified protection level.
Suffix:
<ch>
.
Indicates the protection level.
See also "Protection" on page 536