Remote Control Commands
R&S
®
SMBV100B
897User Manual 1178.4460.02 ─ 03
Parameters:
<State> 0 | 1 | OFF | ON
*RST:  n.a. (factory preset: 1)
Setting parameters: 
<Key> integer
The respective functions are disabled when the protection level
is activated. No password is required for activation of a level.
A password must be entered to deactivate the protection level.
The default password for the first level is 123456.
This protection level is required to unlock internal adjustments
for example.
Example:  To activate protection level:
SYSTem:PROTect1:STATe 1
Internal adjustments or hostname cannot be changed.
To unlock protection level 1:
SYSTem:PROTect1:STATe 0,123456
Internal adjustments are accessible.
Manual operation:  See "Protection Level/Password" on page 537
:SYSTem:SECurity:VOLMode[:STATe] <SecPassWord>, <MmemProtState>
Activates volatile mode, so that no user data can be written to the internal memory per-
manently.
To enable volatile mode, reboot the instrument. Otherwise the change has no effect.
Parameters:
<MmemProtState> 0 | 1 | OFF | ON
*RST:  0
Setting parameters: 
<SecPassWord> string
Current security password
The default password is 123456.
Example: 
SYSTem:SECurity:VOLMode:STATe "123456", 1
SYSTem:REBoot
Manual operation:  See "Volatile Mode" on page 540
:SYSTem:COMMunicate:GPIB:LTERminator <LTerminator>
Sets the terminator recognition for remote control via GPIB interface.
Parameters:
<LTerminator> STANdard | EOI
SYSTem Subsystem