Remote Control Commands
R&S
®
SMBV100B
898User Manual 1178.4460.02 ─ 03
EOI
Recognizes an LF (Line Feed) as the terminator only when it is
sent with the line message EOI (End of Line). This setting is rec-
ommended particularly for binary block transmissions, as binary
blocks may coincidentally contain a characater with value LF
(Line Feed), although it is not determined as a terminator.
STANdard
Recognizes an LF (Line Feed) as the terminator regardless of
whether it is sent with or without EOI.
*RST:  n.a. (factory preset: STANdard)
Example:  See Example "Finding out the used VISA ressource strings"
on page 890.
:SYSTem:COMMunicate:GPIB:RESource?
Queries the visa resource string for remote control via the GPIB interface.
To change the GPIB address, use the command :SYSTem:COMMunicate:GPIB[:
SELF]:ADDRess.
Return values: 
<Resource> string
Example:  See Example "Finding out the used VISA ressource strings"
on page 890.
Usage:  Query only
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <Address>
Sets the GPIB address.
Parameters:
<Address> integer
Range:  0  to  30
*RST:  28
Example:  See Example "Finding out the used VISA ressource strings"
on page 890.
:SYSTem:COMMunicate:HISLip:RESource?
Queries the VISA resource string . This string is used for remote control of the instru-
ment with HiSLIP protocol.
Return values: 
<Resource> string
Example:  See Example "Finding out the used VISA ressource strings"
on page 890.
Usage:  Query only
SYSTem Subsystem